<code>ultracomm</code> is a command line utility for configuring Ultrasonix Tablet systems and acquiring ultrasound data. |
<code>ultracomm</code> is a command line utility for configuring Ultrasonix Tablet systems and acquiring ultrasound data. |
||
− | In the |
+ | In the PhonLab we typically don't run <code>ultracomm</code> directly and instead run the [[ultrasession.py|<code>ultrasession.py</code>]] Python script, which handles a lot of bookkeeping details relevant to a data acquisition session automatically, and which calls <code>ultracomm</code> for you. |
The <code>ultracomm</code> code and releases are available at [[https://github.com/rsprouse/ultracomm github]]. See the 'Releases' section for the latest released version. |
The <code>ultracomm</code> code and releases are available at [[https://github.com/rsprouse/ultracomm github]]. See the 'Releases' section for the latest released version. |
||
Line 32: | Line 32: | ||
The acquisition command can be repeated as many times as desired without re-executing the <code>--init-only</code> command. The imaging parameters will not change unless there is additional user interaction with the ultrasound system. |
The acquisition command can be repeated as many times as desired without re-executing the <code>--init-only</code> command. The imaging parameters will not change unless there is additional user interaction with the ultrasound system. |
||
− | |||
− | == Troubleshooting == |
||
− | |||
− | There is an issue with <code>ultracomm</code> occasionally terminating with an access violation error. This error occurs at program end, after acquisition has completed and datafiles are closed. It is believed to be the result of a bug in the Ulterius library supplied by Ultrasonix and used by <code>ultracomm</code>. Since the error occurs at program end it doesn't seem to result in bad or missed data, but it can result in a Windows error dialog box popping up, which can interfere with the flow of a series of acquisitions in an Opensesame or other experiment. The <code>--av-hack</code> option is used to ignore access violation errors so that they do not result in popup messages. |
Latest revision as of 10:01, 26 April 2018
ultracomm
is a command line utility for configuring Ultrasonix Tablet systems and acquiring ultrasound data.
In the PhonLab we typically don't run ultracomm
directly and instead run the ultrasession.py
Python script, which handles a lot of bookkeeping details relevant to a data acquisition session automatically, and which calls ultracomm
for you.
The ultracomm
code and releases are available at [github]. See the 'Releases' section for the latest released version.
Usage
ultracomm
accepts a number of parameters that are used to connect to and configure the Ultrasonix system. These parameters can be included on the command line or aggregated in a parameter file that is included with the --params
option.
ultracomm --params /path/to/params.cfg
The --address
and --output
parameters are required unless an information-only option was invoked (e.g. --help
or --version
). --address
specifies the IP address of the Ultrasonix system, and --output
names an output file containing a series of ultrasound images.
For a listing of available parameters, see [params.cfg
] in the ultracomm
repository.
For all available options, use:
ultracomm --help
A typical acquistion
A typical acquisition with ultracomm
proceeds by first initializing the ultrasound system and then acquiring data. The first call to ultracomm
is non-interactive. It sets the ultrasound system to use the specified imaging parameters, then freezes the system to the non-imaging mode:
ultracomm --params /path/to/params.cfg --init-only
The second call to ultracomm starts acquiring data and saves to the output file specified by --output
.
ultracomm --params /path/to/params.cfg --output /path/to/output.bpr
This command is interactive and prompts the user to press Enter
to terminate the acquisition. ultracomm
will confirm that the imaging parameters are correct before performing the acquisition.
The acquisition command can be repeated as many times as desired without re-executing the --init-only
command. The imaging parameters will not change unless there is additional user interaction with the ultrasound system.